Manufacturing of Continuous Glucose Monitoring (CGM) Sensors
Technology & Innovation

Manufacturing of Continuous Glucose Monitoring (CGM) Sensors

How continuous glucose monitoring (CGM) sensors are manufactured, including the electrochemical and optical sensing technologies they use, sterilization and factory calibration. It also explains what that manufacturing change means for people who use fingerstick meters and test strips.

Diabetic Supply Rescue Editorial Team
6 min read
Last updated: October 13, 2026
Advanced
12 glossary terms
9 citations

Manufacturing of Continuous Glucose Monitoring (CGM) Sensors

A CGM sensor runs on chemistry that test strip users already rely on. An enzyme, glucose oxidase, reacts with glucose and passes electrons to a metal electrode, and the strength of the resulting current tracks how much glucose is present[1]. The hard part of manufacturing is making thousands of sensors respond to glucose almost identically. Once a company can do that, it can calibrate sensors at the factory instead of asking you to calibrate them with fingerstick readings from your meter[2]. That change explains most of why fingersticks went from a routine CGM requirement to a backup check. (For who makes which sensors, see our guide to major US CGM manufacturers and product lines.)

What a sensor is made of

A test strip gives a measurement of fingerstick blood at a single point in time[1]. A sensor has to keep measuring for days in interstitial fluid, the fluid that surrounds cells just under the skin. The first FreeStyle Libre sensor, for example, combined a thin sensor tail inserted under the skin of the upper arm, an outer casing holding the electronics, and a medical-grade adhesive layer[1].

Two parts decide how an electrochemical sensor responds to glucose, Abbott engineers wrote in 2017: the sensing layer, which holds the enzyme on the working electrode, and a membrane over it that limits how fast glucose reaches the enzyme[2]. Libre sensors use Abbott's Wired Enzyme chemistry, in which an osmium compound (a mediator) shuttles electrons from the enzyme to the electrode[2]. Here's what surprised us: the sensing area is smaller than 1 square millimeter, and the membrane is typically thinner than 100 micrometers. Keeping both nearly identical from sensor to sensor takes high-precision coating equipment[2].

Not every sensor is electrochemical. The implanted Eversense 365 uses a fluorescent indicator molecule grafted onto the sensor's surface. Glucose binds to it reversibly, the glow brightens, and the sensor's optical system measures the change[3].

Made sterile, released lot by lot

The parts that go under the skin ship sterile. FDA's summary for the Dexcom STS, approved in 2006, lists the sensor as a sterile device but not the transmitter, which doesn't touch the skin[4]. Dexcom validated electron-beam sterilization at a minimum dose of 19.3 kGy, with a sterility assurance level of 10⁻⁶, meaning a one-in-a-million chance of a non-sterile unit[4]. For the first FreeStyle Libre, Abbott validated electron-beam sterilization of the sensor pack at a target dose of 25 kGy, reaching the same assurance level[1].

Factory-calibrated sensors are made in lots, and each lot is tested before release[2]. A recall Dexcom started on May 27, 2026 shows what can happen to a lot that never should have left. Certain Dexcom G7 lots marked as scrap were stolen during destruction and later sold to pharmacies and distributors[5]. According to FDA's recall record, one lot had been scrapped because its sensors weren't properly sterilized, and the other because of a higher risk of missing or inaccurate readings[5].

How calibration moved from your meter to the factory

Early sensors drifted, meaning their response to glucose changed over the days of wear, so users had to recalibrate them. After a 2-hour start-up, the Dexcom STS asked for two fingerstick readings from a OneTouch Ultra meter, uploaded by cable; it worked with no other meter[4]. Calibration had to be updated at least every 12 hours, and the labeling said readings had to be confirmed with a blood glucose meter before any therapy change[4]. As late as 2017, Abbott's engineers wrote that drift was the reason every CGM on the market except FreeStyle Libre still required fingerstick calibrations, typically twice a day[2].

Factory calibration, as those engineers describe it, has four manufacturing steps[2]:

  • Make sensor lots with low sensor-to-sensor variability.
  • Pull a sample of sensors from each lot and measure their response to known glucose solutions in the lab.
  • Convert the lot's glucose sensitivity into a sensor code.
  • Show that the sensitivity doesn't change over the sensor's shelf life.

The code can be programmed into the sensor's electronics so nobody has to type it in[2]. If you've read about calibration codes on [test strips](/diabetic-test-strips/calibration-codes-vs-no-code-technology/), the idea will sound familiar: one code stands for a whole manufacturing lot.

User calibration had its own weak spot. If a sensor's signal happened to be briefly off when a meter reading was entered, for example because of an interfering substance, the calculated sensitivity came out wrong and the readings after it stayed falsely high or low[2].

The FDA approved the first FreeStyle Libre in 2017 as a factory-calibrated system that neither required nor accepted calibration with blood glucose values[1]. The Dexcom G6, authorized in 2018, came with a calibration code entered at sensor start. Skipping the code meant two fingerstick calibrations at startup and every 24 hours after that[6]. Dexcom stopped manufacturing the G6 after July 1, 2026, though pharmacies may have stock for a while[7]. FDA's decision summary for the newer Dexcom G7 15 Day sensor describes it as factory calibrated, with an optional fingerstick calibration if the user wants one[8].

Where meters and strips still fit

Factory calibration removed the routine fingerstick, not the meter. The first FreeStyle Libre reader had a built-in blood glucose meter that used FreeStyle Precision Neo test strips[1]. In Abbott's studies, readings from that strip port were independent of the sensor and didn't change its readings[2]. As of October 2026, Abbott directs prescribers to its FreeStyle Libre 3 Plus and 2 Plus sensors, and the earlier Libre 2 and Libre 3 sensors have been discontinued[9].

Labeling still names times for a meter check. FDA's summary of the first Libre says users were told to test with a meter when their symptoms didn't match the sensor, and to use the sensor alongside a meter until they were comfortable with it[1]. The G6 labeling listed times not to rely on the sensor for treatment decisions, including the first two hours of warm-up, when a meter was to be used instead, and any time no reading or trend arrow appeared[6]. Which checks apply to you, and what to do with the result, is a decision for your care team.

The implanted Eversense 365 still depends on strips by design. It needs 4 fingerstick calibrations at the start, then one a day through day 13, then one a week from day 14 on[3].

FDA reviewers flagged a trade-off when they approved the first factory-calibrated sensor. With no required daily meter checks, some users get fewer chances to learn how a particular sensor performs against their meter[1]. Better manufacturing made sensors more consistent, and it also took away the daily reason to pick up the meter.

References

  1. FDA Summary of Safety and Effectiveness Data, P160030: FreeStyle Libre Flash Glucose Monitoring SystemSource
  2. Hoss U, Budiman ES. Factory-Calibrated Continuous Glucose Sensors: The Science Behind the Technology. Diabetes Technology & Therapeutics 2017;19(S2):S44-S50Source
  3. FDA 510(k) Decision Summary K241335: Eversense 365 Continuous Glucose Monitoring (CGM) SystemSource
  4. FDA Summary of Safety and Effectiveness Data, P050012: DexCom STS Continuous Glucose Monitoring SystemSource
  5. FDA Medical Device Recall Z-2574-2026 (Class 2): Dexcom G7 Continuous Glucose Monitoring SystemSource
  6. FDA De Novo Decision Summary DEN170088: Dexcom G6 Continuous Glucose Monitoring SystemSource
  7. Dexcom G6 transitionSource
  8. FDA 510(k) Decision Summary K243214: Dexcom G7 15 Day Continuous Glucose Monitoring SystemSource
  9. Transition to FreeStyle Libre 3 Plus or 2 Plus (FreeStyle Libre Providers)Source