CGM Glossary - Continuous Glucose Monitoring Terms | Diabetic Supply Rescue
Reference Guide

CGM Glossary

A comprehensive reference of continuous glucose monitoring terminology. 102 terms covering sensor technology, accuracy metrics, AID systems, and diabetes management.

5
510(k) Clearance
FDA regulatory pathway for medical devices that are substantially equivalent to existing approved devices. Most CGM improvements use this pathway.
FDAPMADe Novo
A
AID (Automated Insulin Delivery)
Systems that automatically adjust insulin pump delivery based on CGM data. Also known as closed-loop or artificial pancreas systems. Examples include Tandem Control-IQ, Medtronic 780G, and Omnipod 5.
Closed-LoopInsulin PumpAlgorithm
AGP (Ambulatory Glucose Profile)
A standardized CGM report showing glucose patterns over time. Displays median glucose with percentile bands, helping identify trends and problem areas.
CGM DataGlucose PatternsTime in Range
Alarm Fatigue
Desensitization to CGM alerts due to frequent alarms, leading users to ignore or disable warnings. Can be dangerous if genuine alerts are missed.
Alert SettingsHigh/Low AlertsUser Experience
Acetaminophen Interference
Older CGM sensors gave falsely high readings with acetaminophen (Tylenol). Modern iCGM systems like G6/G7 and Libre 3 have resolved this interference.
InterferenceAccuracyMedication
Auto Mode
The automated mode in AID systems where the algorithm controls basal insulin delivery based on CGM readings. Users may need to enter Manual Mode for certain situations.
AIDAlgorithmBasal
A1C (HbA1c)
A blood test measuring average blood glucose over 2-3 months by detecting glycated hemoglobin. Target is typically below 7% for most adults with diabetes.
GMIGlucose ControlAverage Glucose
B
Biofouling
The accumulation of proteins, cells, and tissue around an implanted sensor that can reduce accuracy over time. A major challenge for long-term CGM sensors and implantables.
Foreign Body ResponseSensor DriftBiocompatibility
Beta Bionics iLet
A unique AID system that only requires the user's weight for setup, automatically adjusting all insulin dosing. FDA approved 2023.
AIDBionic PancreasAutomatic
Basal Rate
The continuous background insulin delivery by a pump, expressed in units per hour. AID systems automatically adjust basal based on CGM readings.
Insulin PumpAIDBackground Insulin
Bolus
A dose of insulin given to cover food intake or correct high blood glucose. In hybrid closed-loop systems, users still enter meal boluses manually.
Meal BolusCorrection BolusICR
Bluetooth Low Energy (BLE)
The wireless protocol used by most CGMs to transmit data to receivers and phones. Enables low-power continuous communication.
TransmitterConnectivityWireless
C
CGM (Continuous Glucose Monitor)
A device that continuously measures glucose levels in interstitial fluid through a small sensor inserted under the skin, providing real-time readings and trends throughout the day and night.
Interstitial FluidSensorTransmitter
Closed-Loop System
An automated diabetes management system that combines CGM, insulin pump, and control algorithm to automatically adjust insulin delivery. Minimizes user intervention required.
AIDHybrid Closed-LoopAlgorithm
Control Algorithm
The mathematical software that determines insulin dosing in AID systems. Common approaches include PID (Proportional-Integral-Derivative) and MPC (Model Predictive Control).
AIDPIDMPC
CV (Coefficient of Variation)
A measure of glucose variability calculated as standard deviation divided by mean glucose. CV below 36% indicates stable glucose control.
Glycemic VariabilityStandard DeviationGlucose Control
CE Mark
European conformity certification indicating a medical device meets EU safety and performance requirements. Required for CGM sale in Europe.
FDARegulatoryEurope
Clarity
Dexcom's cloud-based software platform for viewing and analyzing CGM data. Generates AGP reports and provides trend analysis and pattern detection.
DexcomAGPData Analysis
Compression Low
A falsely low CGM reading caused by lying on the sensor, compressing blood flow and glucose transport to the sensor area. Common during sleep.
PISAFalse LowSensor Placement
Clarke Error Grid
A graphical analysis tool for assessing CGM accuracy by plotting CGM vs reference values and categorizing results by clinical significance.
MARDAccuracyConsensus Error Grid
Consensus Error Grid
Updated accuracy assessment tool developed by experts, replacing the Clarke Error Grid. Features five risk zones for evaluating CGM performance.
Clarke Error GridMARDAccuracy
D
DIY Loop
Open-source automated insulin delivery systems created by the diabetes community, including Loop, OpenAPS, and AndroidAPS. Not FDA-approved but used by thousands worldwide.
AIDOpen-SourceAlgorithm
Dexcom
A leading CGM manufacturer based in San Diego. Products include G6, G7, and the integrated Stelo system. Known for accuracy and AID integration.
G6G7iCGM
Dexcom G7
Dexcom's current flagship CGM with 10-day sensor life, 30-minute warmup, all-in-one disposable design, and 8.2% MARD. Approved for ages 2+ and pregnancy.
DexcomiCGMAID
Dexcom G6
Previous generation Dexcom CGM with 10-day sensor life, separate transmitter, and factory calibration. Still widely used and compatible with most AID systems.
DexcomG7Transmitter
Data Sharing
CGM features allowing real-time glucose data to be viewed by caregivers, family members, or healthcare providers remotely via apps like Dexcom Follow or LibreLinkUp.
Remote MonitoringCaregiverMobile App
Dual-Hormone AID
Advanced closed-loop systems that deliver both insulin and glucagon to prevent hypoglycemia more effectively. In clinical trials but not yet commercially available.
AIDGlucagonClosed-Loop
E
Electrochemical Sensing
The method used by most CGMs to measure glucose. An enzyme reaction with glucose produces electrons that create a measurable electrical current proportional to glucose concentration.
Glucose OxidaseElectrodeAmperometric
Eversense
Senseonics' implantable CGM system with a sensor inserted under the skin for up to 6 months. Uses fluorescence technology and requires on-body transmitter.
Implantable CGMSenseonicsLong-term
Early Termination
When a CGM sensor stops working before its intended wear period ends. May be caused by sensor failure, skin reaction, or physical damage.
Sensor LifeSensor ErrorReplacement
F
Factory Calibration
A calibration method where CGM sensors are calibrated during manufacturing, eliminating the need for users to perform finger stick calibrations. Used in Dexcom G6/G7 and FreeStyle Libre systems.
CalibrationUser CalibrationMARD
Full Closed-Loop
An advanced AID system that automates both basal and meal insulin delivery without requiring meal announcements. Still in development and clinical trials for most use cases.
Closed-LoopHybrid Closed-LoopAlgorithm
Fluorescence Sensing
An optical glucose sensing method used in Eversense implantable CGM. Uses fluorescent molecules that change light emission based on glucose concentration in interstitial fluid.
Implantable CGMOpticalEversense
Foreign Body Response (FBR)
The body's natural immune and wound healing response to an implanted sensor. Can create a fibrous capsule that reduces glucose diffusion to the sensor, affecting accuracy.
BiofoulingSensorBiocompatibility
FreeStyle Libre
Abbott's flash glucose monitoring system that requires scanning to see readings. The Libre 3 provides real-time continuous monitoring. Known for 14-day wear and low cost.
AbbottFlash GlucoseLibre 3
FreeStyle Libre 3
Abbott's smallest CGM sensor with real-time glucose monitoring, optional alerts, and 14-day wear. Features 7.9% MARD and no scanning required.
FreeStyle LibreAbbottReal-time CGM
Flash Glucose Monitoring
A type of CGM that stores continuous glucose data but requires users to scan the sensor with a reader or phone to see readings, rather than continuously transmitting.
CGMScanningFreeStyle Libre
G
GMI (Glucose Management Indicator)
An estimate of A1C derived from CGM data, based on the relationship between average glucose and A1C. Formerly called eA1C (estimated A1C).
A1CAverage GlucoseCGM
Glycemic Variability
The degree of glucose fluctuation over time, measured by metrics like CV (coefficient of variation) and standard deviation. CGM reveals variability that finger sticks cannot capture.
Time in RangeStandard DeviationCV
Glucose Oxidase (GOx)
The enzyme most commonly used in CGM sensors to catalyze the glucose reaction that produces a measurable electrical signal. Highly specific for glucose but sensitive to oxygen levels.
EnzymeElectrochemicalGDH
Glucose Dehydrogenase (GDH)
An alternative enzyme used in some glucose sensors. Not affected by oxygen levels but may react with other sugars like maltose depending on the cofactor used.
Glucose OxidaseEnzymeSpecificity
Guardian 4
Medtronic's latest CGM sensor designed for use with the 780G insulin pump system. Features 7-day wear and integration with Medtronic's SmartGuard technology.
Medtronic780GSmartGuard
Glucose Target
The blood glucose level that AID systems aim to achieve, typically 100-120 mg/dL. Some systems allow user-adjustable targets for activities or sleep.
Target RangeAIDSettings
Gestational Diabetes
Diabetes diagnosed during pregnancy. CGM provides detailed glucose data to help maintain tight control for maternal and fetal health.
PregnancyGlucose ControlFetal Health
H
Hybrid Closed-Loop
A type of AID system that automates basal insulin delivery but still requires user input for meal boluses. Current commercial systems like Control-IQ, 780G, and Omnipod 5 are hybrid closed-loop.
Closed-LoopAIDMeal Bolus
Hydroxyurea Interference
A medication used in sickle cell disease that can cause falsely elevated CGM readings. One of few remaining interferences in modern CGM systems.
InterferenceAccuracyMedication
Hypoglycemia
Low blood glucose, typically defined as below 70 mg/dL, or severe below 54 mg/dL. CGM predictive alerts help prevent hypoglycemic episodes.
Low GlucoseUrgent LowSafety
Hyperglycemia
High blood glucose, typically defined as above 180 mg/dL. CGM helps identify hyperglycemia patterns and triggers for treatment adjustments.
High GlucoseTime Above RangeTreatment
I
Interstitial Fluid (ISF)
The fluid between cells in body tissue where CGM sensors measure glucose. ISF glucose lags behind blood glucose by approximately 5-15 minutes due to the time required for glucose to diffuse from capillaries.
CGMLag TimeSensor
iCGM
Integrated Continuous Glucose Monitor - an FDA classification for CGM systems that meet specific accuracy and reliability standards allowing integration with other devices like insulin pumps.
FDAAIDInteroperability
Interoperability
The ability of CGM systems to communicate with and be used by other medical devices, particularly insulin pumps and AID systems from different manufacturers.
iCGMAIDConnectivity
Implantable CGM
CGM sensors that are surgically placed under the skin for extended wear periods (3-6 months). Senseonics Eversense is currently the only available implantable CGM.
EversenseFluorescenceLong-term
IBOA (Isobornyl Acrylate)
A chemical component in some CGM adhesives that can cause allergic contact dermatitis. Abbott and Dexcom have modified formulations to address this issue.
Skin ReactionAdhesiveAllergy
Insulin Pump
A small device that delivers rapid-acting insulin continuously through tubing or a patch, controlled by CGM data in AID systems.
AIDBasalBolus
In Silico
Computer simulation studies used to test CGM algorithms and AID systems before human trials. Regulatory agencies accept in silico data for some submissions.
AlgorithmTestingFDA
L
Lag Time
The delay between blood glucose changes and corresponding CGM readings, typically 5-15 minutes. Caused by glucose diffusion from blood to interstitial fluid and sensor processing time.
Interstitial FluidAccuracyPhysiologic Lag
Lingo
Abbott's over-the-counter biosensor system for metabolic health insights, based on Libre technology. Marketed for weight management and energy optimization.
AbbottOTCWellness
LibreView
Abbott's cloud-based platform for viewing and sharing FreeStyle Libre CGM data with healthcare providers. Generates standardized reports.
FreeStyle LibreAbbottData Sharing
M
MARD (Mean Absolute Relative Difference)
The standard metric for evaluating CGM accuracy, calculated as the average percentage difference between CGM readings and reference blood glucose values. Lower MARD indicates higher accuracy. Current CGMs achieve MARD of 8-12%.
AccuracyCalibrationClarke Error Grid
MPC (Model Predictive Control)
An advanced control algorithm that predicts future glucose levels and optimizes insulin delivery accordingly. Used in systems like Control-IQ and 780G.
Control AlgorithmPIDAID
Membrane
The outer coating of a CGM sensor that controls glucose flux and protects the sensor chemistry. Also serves to improve biocompatibility and reduce interference.
SensorBiocompatibilityBiofouling
MRI Safety
CGM sensors must be removed before MRI scans due to metal components and potential for burns or sensor damage. Some newer sensors have improved MRI labeling.
SafetyMetalMedical Procedure
Microneedle CGM
Emerging CGM technology using tiny needles that penetrate only the outermost skin layer, potentially offering minimally-invasive monitoring with less foreign body response.
Minimally InvasiveSensorFuture Technology
Multi-Analyte Sensing
Future CGM technology that measures multiple biomarkers (lactate, ketones, alcohol) in addition to glucose, providing more comprehensive metabolic insights.
CGMBiomarkersFuture Technology
Medtronic 780G
Medtronic's advanced hybrid closed-loop system with 5-minute auto-corrections and SmartGuard technology. Uses Guardian 4 sensors.
AIDGuardianSmartGuard
N
Non-Adjunctive
FDA designation allowing CGM readings to be used for diabetes treatment decisions without confirmatory finger stick tests. Indicates high accuracy and reliability.
Dosing ClaimAccuracyiCGM
Non-Invasive Glucose Monitoring
Experimental technologies attempting to measure glucose without breaking the skin. Approaches include optical, RF, and bioimpedance methods. None are FDA-cleared for medical use.
Optical SensingSpectroscopyWearable
Near-Infrared Spectroscopy (NIR)
An optical technique that shines near-infrared light through tissue to measure glucose based on light absorption patterns. Challenged by water and tissue interference.
OpticalNon-InvasiveSpectroscopy
NFC (Near Field Communication)
Wireless technology used for CGM scanning and activation. FreeStyle Libre uses NFC for scan-to-read and sensor activation.
Flash GlucoseScanningFreeStyle Libre
O
Optical Glucose Sensing
Non-invasive glucose measurement using light, including near-infrared spectroscopy, Raman spectroscopy, and photoacoustic methods. Major technical challenges remain unsolved.
Non-InvasiveSpectroscopyNIR
Omnipod 5
Insulet's tubeless, waterproof patch pump with integrated Dexcom G6 AID capability. The first tubeless AID system approved in the US.
AIDPatch PumpDexcom
P
PID Control
Proportional-Integral-Derivative control algorithm used in some AID systems. Reacts to current glucose levels, recent history, and rate of change to adjust insulin delivery.
Control AlgorithmMPCAID
Physiologic Lag
The inherent delay between blood glucose and interstitial fluid glucose, typically 5-10 minutes. A fundamental limitation of CGM that cannot be eliminated.
Lag TimeInterstitial FluidISF
Predictive Alerts
CGM alarms that warn users before glucose reaches high or low thresholds by predicting future glucose based on current trends. Helps prevent hypo/hyperglycemia.
AlertTrend ArrowAlgorithm
PMA (Premarket Approval)
FDA's most stringent regulatory pathway for high-risk medical devices. AID systems typically require PMA approval.
FDA510(k)Regulatory
PISA (Pressure-Induced Sensor Attenuation)
False low glucose readings caused by pressure on the CGM sensor, often during sleep. Results in inaccurate low alarms and can disrupt sleep quality.
Compression LowSensor PlacementAccuracy
R
Receiver
A dedicated device that displays CGM glucose readings and trends. Many users now use smartphone apps instead of dedicated receivers.
TransmitterDisplayMobile App
Rate of Change
How quickly glucose is rising or falling, typically measured in mg/dL per minute. Rates above 2 mg/dL/min are considered rapid changes requiring attention.
Trend ArrowAlertsPredictive
Real-time CGM (rtCGM)
CGM systems that automatically transmit glucose readings to a display device at regular intervals (typically every 1-5 minutes) without requiring user scanning.
Flash GlucoseContinuous MonitoringAlerts
Remote Monitoring
The ability for caregivers or healthcare providers to view CGM data in real-time from a distance. Particularly valuable for parents of children with diabetes.
Data SharingFollow AppCaregiver
Reference Measurement
Laboratory-quality blood glucose measurement used to assess CGM accuracy. Yellow Springs Instrument (YSI) analyzers are the gold standard.
MARDAccuracyYSI
S
Sensor
The component of a CGM system that is inserted under the skin to measure glucose levels. Contains electrochemical or optical elements that detect glucose in interstitial fluid.
CGMTransmitterElectrode
Sensor Life
The approved duration a CGM sensor can be worn before replacement. Varies from 7 days (Guardian) to 14-15 days (FreeStyle Libre/Dexcom G7) to 6 months (Eversense implantable).
SensorWear TimeReplacement
Sensor Drift
Gradual change in CGM accuracy over the sensor's wear period, typically showing less accurate readings as the sensor ages due to biofouling and enzyme degradation.
AccuracyBiofoulingCalibration
Stelo
Dexcom's over-the-counter CGM designed for people without diabetes. Provides glucose insights without prescription, targeting the wellness market.
DexcomOTC CGMWellness
Signal Loss
Temporary interruption of CGM data transmission, often due to sensor-transmitter-receiver distance, Bluetooth interference, or sensor issues.
BluetoothTransmitterConnectivity
Sensor Error
A CGM message indicating the sensor cannot provide accurate readings, often requiring waiting or replacement. May be caused by biofouling, bleeding, or sensor malfunction.
SensorAccuracyTroubleshooting
Skin Reaction
Allergic or irritant response to CGM sensor adhesive or components. Symptoms include redness, itching, rash, or blistering. May require barrier products or sensor changes.
AdhesiveIBOAAllergy
Smart Contact Lens
Experimental technology to measure glucose in tear fluid using a sensor embedded in a contact lens. Google/Novartis project was discontinued; research continues elsewhere.
Non-InvasiveTear GlucoseWearable
Smartwatch CGM
Proposed technology to measure glucose from a watch-like device without any implanted sensor. Multiple companies claim development but none have achieved clinical accuracy.
Non-InvasiveWearableApple Watch
Suspend Before Low
An AID feature that automatically stops insulin delivery when the CGM predicts glucose will drop below a set threshold, helping to prevent hypoglycemia.
Predictive Low Glucose SuspendSafetyHypoglycemia
T
Transmitter
The electronic component that wirelessly sends glucose data from the sensor to a receiver, smartphone, or insulin pump. May be reusable or integrated into disposable sensors.
SensorReceiverBluetooth
Time in Range (TIR)
The percentage of time glucose levels stay within a target range, typically 70-180 mg/dL. A TIR of 70% or higher is a common goal. CGM enables accurate TIR measurement.
CGMTarget RangeGlycemic Control
Time Below Range (TBR)
The percentage of time glucose is below the target range (typically below 70 mg/dL). Goal is less than 4% TBR, with less than 1% below 54 mg/dL.
Time in RangeHypoglycemiaTarget Range
Time Above Range (TAR)
The percentage of time glucose is above the target range (typically above 180 mg/dL). Goal is less than 25% TAR, with less than 5% above 250 mg/dL.
Time in RangeHyperglycemiaTarget Range
Trend Arrow
Visual indicator showing the direction and rate of glucose change. Arrows indicate if glucose is stable, slowly rising/falling, or rapidly changing. Critical for dosing decisions.
CGMRate of ChangeDisplay
Tandem Control-IQ
Tandem's hybrid closed-loop system using Dexcom G6/G7 CGM. Automatically adjusts basal and can deliver automatic correction boluses. FDA approved 2020.
AIDDexcomHybrid Closed-Loop
Type 1 Diabetes
An autoimmune condition where the pancreas produces little or no insulin. People with Type 1 require insulin therapy and benefit significantly from CGM and AID.
InsulinAIDAutoimmune
Type 2 Diabetes
A metabolic condition characterized by insulin resistance. CGM is increasingly used in Type 2 for medication optimization and behavior change.
Insulin ResistanceMedicationLifestyle
U
User Calibration
Traditional calibration method requiring users to enter finger stick blood glucose readings into the CGM system to maintain accuracy. Typically required 2-4 times per day with older systems.
Factory CalibrationFinger StickAccuracy
Urgent Low Soon Alert
A predictive alarm that warns when glucose is predicted to drop below 55 mg/dL within 20 minutes, allowing time to treat and prevent severe hypoglycemia.
Predictive AlertsHypoglycemiaSafety
W
Warm-up Period
The time after sensor insertion before the CGM provides readings, allowing the sensor to stabilize. Ranges from 30 minutes (Dexcom G7) to 2 hours (Guardian Connect).
SensorInsertionAccuracy

Have Unused CGM Supplies?

Don't let your extra Dexcom, FreeStyle Libre, or other CGM supplies go to waste. Turn them into cash with our easy sell program.

Get a Free Quote